Skip to content
Multiskan-FC
Features:
- There should be a provision of broad wavelength range of 340 – 850 nm for a wide variety of research and routine applications.
- The system should measure fast and accurate of 96- well plates for varying throughput requirements
- There should be Shaker for critical assays
- There should be Ease of use through the large color screen, visual internal software and a variety of language versions
- The system must be operating with Visual and logical SkanIt Software for comprehensive instrument control and data handling
Technical Specification:
- Light source : Quartz-halogen lamp 6 V / 10 W
- Wavelength range: 340 – 850 nm
- Filters: 8-position filter wheel, the instrument is delivered with the following standard filters installed: 405 nm; 450 nm; and 620 nm. Additional filters can be ordered separately.
- Half-bandwidth of filters: 3 – 9 nm
- Read-out range: 0 – 6 Abs
- Linearity (405 nm): 0 – 3 Abs, ± 2%, 96-well plate, fast mode
0 – 4 Abs, ± 2%, 96-well plate,
normal mode
- Resolution : 0.001 Abs
- Accuracy (405 nm): ± 1% (0.3 – 3 Abs), ± 2% (3 – 4 Abs)
- Precision (405 nm) : CV ≤ 0.2% (0.3 – 3 Abs), CV ≤ 1.0% (3 – 4 Abs)
- Measurement speed: 6 s, 96-well plate, fast mode
12 s, 96-well plate, normal mode
- Shaking: Linear shaking with three modes: slow, medium and fast
- Robotic compatibility: Yes
- Display: High contrast color display (480 x 272 dots)
- User interface : Internal software or PC control with SkanIt Software
Internal memory (standalone) At least up to 99 assay protocols and 100 test results, 96-well plate
- External printer type : HP PCL5
- Communication :
- USB for computer connection
- USB memory stick position for data export
- USB for external printer
- CE Certified