Monitor -100
Specifications:
- Throughput: 200Test/hr. in Random Access.
- Sample Position: 100 samples at a time.
- ESR Tubes : MONOSED Vacuum/Non-Vacuum Tubes With only 1.28 ml whole blood
- Analysis Time: 30 min/sample.
Print results corrected to 1 hr. in mm westergreen.
- Measuring Method: Infra Red Barrier.
- Temp. Correction: Automatic compensation referred to 18 c (Manley)
- QC software : Built-in. Printing of the Levey Jennings Charts
- Display: Graphic LCD with backlight.
Display 81 print out The Sedimentation Curve
- Results : In Westerngreen mm/ hr. (by interpolation)
- Result Resolution: +‑ 1mm.
- Keyboard: 15 keys.
- Patient ID: By Bar Code Reader or Keyboard.
- Operating Condition : RT from 15-32 c, humidity: 45%-85%
- Printer Interface: RS 232 serial output.
- CCD Scanner Interface: TRserial input.
- Instrument Size: HxWxD (185x290x445) mm
- Power: 100-240V AC, 50/60 Hz
- Weight: 7.2 kg
- Conformity: CE Approved.
IVD Directive 98/79/EEC.
EMC Directive 89/336/EEC.